G-2020
Technical Specifications:
Function | pH OR mV & temperature simultaneously |
Display | 16 character x 2 lines Alphanumeric BL LCD |
Power supply | 230 V ± 10% 50 Hz AC |
pH | |
Range | 0 to 14 pH |
Resolution | 0.01 pH |
Accuracy | ± 0.01 pH ± 1 digit |
Slope adjustment | 80% to 120% By default 100% |
Temp.compensation | 0 to 100 ˚C Auto with temp probe / Manual with keys |
Auto buffer recognition | 4.00 pH, 7.00 pH,9.20 pH |
Calibration | Auto as manual, 3 point / 2 point (Programmable) |
mV | |
Range | -1999 mV to +1999 mV |
Resolution | 1 mV |
Accuracy | ± 0.1% ±1 digit |
Calibration | Two point manual |
Temperature | |
Range | -50.0 ˚C to 200 ˚C |
Resolution | 0.1 ˚C |
Accuracy | ± 0.1% of range ±1 digit |
Calibration | Two point manual |
Sensors | Pt100 probe |
Data entry | By six soft touch keys with audible sound |
Storage | 1000 samples |
Real time display | 24 Hour mode with Date |
Interface | RS-232, Centronics & computer |
Print format | All Data & Date Wise (Programmable) |
Digital output | Serial RS-232 Parallel Centronics (Programmable) |
Error Indication | pH ERROR,ORP ERROR,CAL ERROR & T-ERROR |
Accessories | pH electrode Instruction Manual Buffer Solutions Electrode Stand Temp Probe |
Appearance and specification subject to change due to continuous development.